|
The
nitrogen tank (size D) should be replaced when the
high pressure gauge reads 100 psi or less. To replace
nitrogen tank disconnect the water chamber and turn
off the nitrogen tank valve with the nitrogen wrench.
Then set pump on its end with the nitrogen tank valve
pointing down. Loosen band screw around tank and loosen
chrome yoke "T" handle to free the tank.
Remove tank and replace with a full tank. The new
tank will have a new plastic seal attached to the
medical post. Do not use it unless the old seal, which
will be in the medical yoke is damaged. Never use
two seals. Only nitrogen (N2) should be used. Position
the tank's medical post into yoke so the two pins
and the inlet in the yoke fit into the holes in the
medical post. Tighten chrome "T" handle
very tight. Tighten band clamp only slightly. Set
pump down on it's feet and open nitrogen tank valve
with nitrogen wrench slowly. The high pressure gauge
(0 to 3000psi) should read 1800 to 1900psi. Adjust
the low pressure (0 to 30psi)to 15psi with the brass
"T" handle on the top of the regulator,
which is on the top right side of the cabinet. Now
turn off the nitrogen tank with the nitrogen wrench
and turn the brass "T" handle counter clockwise
and remove it from the regulator. Note on paper the
high pressure (typically 1800 to 1900psi) and also
note on paper the low pressure (typically 12 to 15psi).
Check the high and low pressure in 12 to 24 hours.
The high pressure should have not dropped more than
200psi and the low pressure should not have dropped
more than 2psi. Now that you have established that
you have leak free system you may open the nitrogen
tank valve all the way counter clockwise and leave
it open until the high pressure gauge reads 100psi
or less, typically 3 to 6 months.
If
the pump does not pass the pressure-leak test you
must find where it is leaking. To assist you in locating
the leak mix water and liquid dishwashing detergent
(10 to 1) in a small plastic squeeze bottle. If the
leak is a high-pressure leak apply water-detergent
to the connection between the medical yoke and medical
post, to all high-pressure threaded fittings, including
the high pressure gauge. If on the other hand the
leak is on the low the side, apply water-detergent
to alllow-pressure threaded fittings. A large leak
will show up as big soap bubbles, a very slow leak
will look like foam or shaving cream. All leak tests
should be done with the water chamber out of the system.
If
your pump tests okay, but your nitrogen consumption
is high your leak maybe in the water chamber. To test
this empty the water chamber and replace the lid.
Now connect the nitrogen line (top line) to the water
chamber: Apply 15psi of nitrogen to the water chamber,
while it is completely submersed in water. Note any
bubbles that would indicate a leak.
|